Posted on 1/29/18 at 7:44 pm to Man4others
When the magnet quenches it liquifies air by the magnet. Maybe someone quenched it and finished him off.
Posted on 1/29/18 at 8:48 pm to LEASTBAY
quote:
I used to work in MRI. I have a BS. If you quench the magnet it can actually destroy the building. Look up some YouTube vids of it happening. That magnet is always on. Ain't no 100k fix either.
Damage to the magnet room is kind of rare, it happens when the cryogen vent is improperly installed or is defective in some way which allows the super cooled and expanding helium gas to vent into the room instead of out vent. The sudden increase in room pressure can buckle walls, and blow down ceilings. I was involved in one vent failure, the cost of the room damage, and ramping the magnet was around $250K. In that case the vent failure was caused by a defect in the vent piping and the manufacture eventually paid the bill. Recovery from a quench with no damage, or unusual magnet issues can be done for under a $100K if a manufacturer does it. If you contract it out to a non OEM service provider it can be done for under $50k, usually way under.
This post was edited on 1/29/18 at 8:54 pm
Posted on 1/29/18 at 9:04 pm to LEASTBAY
quote:
When the magnet quenches it liquifies air by the magnet. Maybe someone quenched it and finished him off.
MRI magnets are mostly of superconducting design. They require liquid helium to maintain the very low (-452F) temperature needed for the magnet coils to be superconductive. To quench a magnet a small amount of heat is generated inside the magnet which starts a reaction that rapidly vaporizes the liquid helium inside the magnet into helium gas. When a quantity of liquid is turned to gas the volume increases greatly, rather than letting this expanding gas be trapped in the magnet resulting in a explosion, a elaborate venting system is used to vent the expanding gas outside the magnet room. Under normal circumstances no gas of any kind will escape into the magnet room during a magnet quench. There have been a relatively small number of cases where the vent system has failed during a quench.
I know, TMI!
This post was edited on 1/29/18 at 9:05 pm
Posted on 1/29/18 at 9:06 pm to TigersFan64
How do magnets work again?
Posted on 1/29/18 at 9:55 pm to EA6B
quote:
$50-100 grand if it can be recovered, there is a small chance of permanent damage, then it could be $500k-750K to replace the magnet.
I thought these things were a couple million?
Posted on 1/29/18 at 10:18 pm to BearsFan
quote:
I thought these things were a couple million?
That is for the initial purchase of the entire system, just the physical magnet without the rest of the system is probably in the neighbor hood of $300k, but it depends on if it is new or used and who you buy it from. Then there is the labor and other expenses involved in physically replacing it, many variables involved.
Posted on 1/30/18 at 7:09 am to TigerstuckinMS
quote:
If he was in between the bottle and the magnet, like if his back were to the machine and he was carrying the bottle with both hands across his body, it wouldn't matter if he let it go. If it was pulling the bottle against his center of mass, once the magnet had enough pull on the bottle to break his footing and start pulling him toward the machine, it would be all over in the blink of an eye.
Is it bad that I hope liveleak gets video footage of this?
Posted on 1/30/18 at 7:13 am to Langland
quote:
How did he inhale liquid oxygen?
It's in India.
Ask Dr. Pvnrt
Posted on 1/30/18 at 9:02 am to Breesus
quote:
It's in India.
Ask Dr. Pvnrt
ISWYDT

Posted on 1/30/18 at 9:03 am to TH03
Those signs are there for reason.
Posted on 1/30/18 at 9:04 am to TigersFan64
Stupid, stupid, stupid
A magnet used to pick up cars in a junkyard is 1 Tesla. That MRI machine was prob 1.5 Tesla.
A magnet used to pick up cars in a junkyard is 1 Tesla. That MRI machine was prob 1.5 Tesla.
This post was edited on 1/30/18 at 9:07 am